21 CFR Part 11, explainedElectronic records and signatures the FDA will accept
Part 11 is the FDA rule that lets a regulated company keep its records, and sign them, electronically instead of on paper, as long as the system can prove who did what, when, and that nothing was quietly changed. This guide takes each requirement in turn, then covers what inspectors look for and what to ask of software.
- Issued by
- US FDA, Title 21 of the Code of Federal Regulations, Part 11
- Applies to
- Electronic records that FDA's record rules require (drug GMP in 21 CFR 211, devices, GLP and others), electronic records submitted to FDA, and the signatures on them. Foreign sites supplying the US are inspected against it
- In force
- Published 20 March 1997, effective 20 August 1997; enforcement narrowed by FDA guidance in 2003
In plain words
What 21 CFR Part 11 is about
Part 11 does not decide which records you keep. The predicate rules do, for example 21 CFR Part 211 for drug manufacturing. Part 11 decides when an electronic version of those records, and an electronic signature on them, is as trustworthy as paper and ink.
It has three subparts: general provisions (scope, implementation and definitions), electronic records (controls that keep records authentic and intact, so a signer cannot readily disown a signed record as not genuine) and electronic signatures (how a signature is made, linked to its record and protected).
Which records it covers was set out in FDA's 2003 guidance on its scope and application: a required record kept electronically instead of on paper, or kept both ways when the electronic version is the one relied on for regulated work; electronic records submitted to FDA; and electronic signatures meant to replace handwritten ones. FDA recommends deciding in advance, and writing down, which version of each record you rely on.
The same guidance narrowed enforcement. FDA said it would use enforcement discretion on validation, audit trails, record retention and record copying, and on legacy systems that were operating before 20 August 1997, met the predicate rules then, still meet them, and are documented as fit for use. It still enforces the rest of Part 11 and the predicate rules themselves. In practice inspectors still expect validated systems with audit trails for GMP records, because the predicate rules and FDA's data integrity guidance call for them.
What it asks for
The requirements, clause by clause
- §11.10(a)
Validated systems
Validate the system for accuracy, reliability and consistent intended performance, and for its ability to tell when a record is invalid or has been altered.
- §11.10(b)
Accurate copies
Produce complete, accurate copies in human-readable and electronic form that FDA can inspect, review and copy.
- §11.10(c)
Protected records
Protect records so they can be retrieved accurately and readily throughout the retention period.
- §11.10(d)
Limited access
Limit system access to authorised individuals.
- §11.10(e)
Audit trail
Keep secure, computer-generated, time-stamped audit trails that independently record the date and time of operator entries and actions that create, modify or delete records. Changes must not obscure what was recorded before. The trail is kept at least as long as the records and is available for FDA to review and copy.
- §11.10(f)–(h)
System, authority and device checks
Enforce the permitted sequence of steps, check that the user is authorised for each action (signing, altering a record, using an input or output device) and, where needed, check that the data source is valid.
- §11.10(i)
Trained people
Everyone who develops, maintains or uses the system has the education, training and experience for their tasks.
- §11.10(j)
Accountability policy
Written policies hold people accountable for actions taken under their electronic signatures, to deter falsification.
- §11.10(k)
System documentation
Control the distribution of, access to and use of system documentation, with revision and change control that keeps a time-sequenced record of how the documentation was developed and changed.
- §11.30
Open systems
Where access to a system is not controlled by the people responsible for its records, apply the §11.10 controls as appropriate, plus measures such as document encryption and digital signature standards where needed to keep records authentic, intact and confidential.
- §11.50
What a signature shows
A signed record shows the signer's printed name, the date and time of signing, and the meaning of the signature, such as review, approval or authorship, on screen and on printouts.
- §11.70
Signature linking
Electronic signatures, and handwritten signatures on electronic records, are linked to their records so they cannot be removed, copied or otherwise transferred to falsify another record by ordinary means.
- §11.100
One signature, one person
Each e-signature belongs to one individual and is never reused by or reassigned to anyone else, and identity is verified before it is issued. Before or at first use, the company certifies to FDA, over a handwritten signature, that its e-signatures are the legally binding equivalent of handwritten ones; FDA can ask for more on any one signature.
- §11.200
Two components
Non-biometric signatures use at least two distinct components, such as a user ID and a password. The first signing in a continuous session uses all of them, later ones at least one that only the owner can execute; signings outside one session use all of them. Only the owner uses a signature, and misusing it would take two or more people colluding.
- §11.300
ID and password controls
ID and password combinations are unique, checked and revised periodically, and deauthorised when lost or compromised. Safeguards prevent unauthorised use and report attempts immediately to security and, as appropriate, management. Tokens and cards that carry or generate codes are tested at first and periodically.
Where inspections find gaps
Where sites usually slip
- Shared or generic loginsActions are recorded against "QC1" or "admin", so no entry can be attributed to a person.
- Audit trails off or unreadThe audit trail is disabled, does not cover GMP-relevant data, or is never reviewed.
- Users who can erase their own dataAnalysts or operators with administrator rights can delete or overwrite records they generated.
- Incomplete signaturesSignatures that do not show name, date, time and meaning, or that are not bound to the record they sign.
- Missing paperworkNo e-signature certification filed with FDA, or no written policy making people accountable for their e-signatures.
- Unvalidated or uncontrolled systemsThe system was never validated for its intended use, or it has been changed without change control since.
What to ask of software
Questions for any vendor
At a glance
Part 11 and Annex 11 side by side
Most regulated manufacturers that export work to both. The two cover the same ground from different directions: Part 11 is a short US regulation about records and signatures; Annex 11 is an EU guideline about the whole life of a computerised system.
| 21 CFR Part 11 (US FDA) | EU GMP Annex 11 | |
|---|---|---|
| Legal form | Federal regulation (1997), binding in the US | Guideline in the EU GMP Guide (2011); also used by PIC/S inspectorates |
| What it covers | Electronic records required by FDA rules, and electronic signatures | Every computerised system used in GMP activities, across its lifecycle |
| Risk | Narrowed by FDA's 2003 guidance to a risk-based scope | Risk management across the lifecycle is the first section |
| Audit trail | Computer-generated, time-stamped record of create, modify and delete, kept at least as long as the records (§11.10(e)) | Based on risk; records GMP-relevant changes and deletions with the reason, readable and regularly reviewed (9) |
| Electronic signature | Name, date, time and meaning; linked to the record; two components; a hand-signed certification to FDA (§11.50–11.300) | Same effect as a handwritten signature within the company, permanently linked, with date and time (14) |
| Suppliers | Not addressed directly | Formal agreements; supplier assessment and audits based on risk (3) |
| Keeping it validated | Validation and system documentation control (§11.10(a), (k)) | Change and configuration management, periodic evaluation, incident management (10, 11, 13) |
| Continuity and archive | Records protected and retrievable for the retention period (§11.10(c)) | Backups with tested restore, business continuity and archiving (7, 16, 17) |
| Status | Requirements unchanged since 1997; FDA guidance explains how it applies them | 2011 text in force; a revised draft went to consultation in July 2025 and is not yet final |
Where GAMP Lifetech helps
The modules behind 21 CFR Part 11 records
- ADMINEvery platformSystem AdministratorOne login across every module, with single sign-on and Active Directory integration, password policy set centrally, and locked accounts reset through one controlled route: the identity and password controls of §11.100 and §11.300.See the module
- DMSComplianceOneDocument Management SystemDocuments approved by electronic signature, with version control and full traceability from creation to archive: the signature meaning, date and linking of §11.50 and §11.70.See the module
- TMSComplianceOneTraining Management SystemTraining, assessments and competency records for everyone who uses your systems: the evidence §11.10(i) asks for.See the module
- UAMComplianceOneUser Access ManagementAccess to your other GxP systems and instruments requested, approved, changed and removed through one recorded route: §11.10(d) beyond the GAMP suite.See the module
- QMSComplianceOneQuality Management SystemChange control with impact assessment when a system changes, and deviations and CAPA when something goes wrong.See the module
Every GAMP Lifetech module keeps its own computer-generated audit trail, electronic signatures and role-based access, with single sign-on across the suite through System Administrator.
FAQs
Questions people ask about 21 CFR Part 11
Does 21 CFR Part 11 apply to Indian pharmaceutical manufacturers?
Yes, if you make products for the US market or submit records to FDA. FDA's GMP rules apply to foreign sites that supply the US, and Part 11 applies to the electronic records and signatures those rules require. An Indian plant supplying the US is inspected against it.
Is a printout from a computer system a Part 11 record?
Generally not, when the printout meets the predicate rules and the paper is what people rely on; the computer only produced it. If people rely on the electronic version, FDA may treat that as the record. Part 11 also does not cover paper records that are only sent electronically. And FDA's 2018 data integrity guidance warns that a printout of a dynamic record, such as an FT-IR spectrum that can be reprocessed, is not a true copy of it.
Is software ever "Part 11 certified"?
No. FDA does not certify software, and no vendor can make you compliant. Software can provide the technical controls Part 11 requires; compliance depends on how you validate, configure and use the system, and on your procedures.
What must an electronic signature show under Part 11?
Under §11.50, the signer's printed name, the date and time the signature was executed, and its meaning, such as review, approval, responsibility or authorship. These must appear wherever the record is displayed or printed, and §11.70 requires the signature to be linked so it cannot be copied to another record.
Does Part 11 require audit trail review?
Part 11 requires the audit trail and its retention (§11.10(e)) but sets no review interval. FDA's 2018 data integrity guidance fills the gap: the people who review a record under GMP review the audit trail that captures changes to it as part of that review, at the frequency the GMP rules set for the record, such as before batch release. Where the rules set no frequency, it is set by risk.
Can people share a login?
Not for any action that changes data. FDA's 2018 data integrity guidance says a shared login cannot identify a unique person, so the system would not meet GMP requirements for attributable actions. Shared, read-only accounts that cannot change data or settings are acceptable for viewing data.
Is a user ID and password the same as two-factor authentication?
No. §11.200 asks for two distinct identification components, such as an ID code and a password, to execute a non-biometric e-signature. A user ID is not secret, so in security terms that is one factor. Part 11 does not require multi-factor authentication; many companies add it for remote or privileged access as their own security control.
What is the FDA certification for electronic signatures?
Under §11.100(c), before or at the time it first uses electronic signatures, a company certifies to FDA that the electronic signatures in its system are intended to be the legally binding equivalent of traditional handwritten signatures. The certification is signed by hand, and FDA can ask for further certification or testimony that a specific electronic signature is binding.
Can we mix paper and electronic records?
Yes, but each part has to meet its own rules, and the link between a handwritten signature and the electronic record it applies to has to be controlled. Hybrid systems are harder to keep in control, and they are a common source of findings.
Keep reading
Next, read these
- EU GMP Annex 11How every computerised system used in GMP is validated, secured and kept in control across its life.
European Commission
EUAnnex 11
16 requirements7 min read - ALCOA++The ten qualities that make a GxP record trustworthy, on paper and in software.
FDA, MHRA, PIC/S, EMA
GlobalALCOA++
5 requirements5 min read - GAMP 5ISPE's risk-based approach to validating computerised systems, and its software categories.
ISPE
GlobalGAMP 5
8 requirements6 min read
From the library
Official sources
- 21 CFR Part 11 (eCFR)
- FDA guidance: Part 11 scope and application (2003)
- FDA guidance: Data integrity and compliance with drug CGMP (2018)
- 21 CFR Part 11 in the GxP glossary
This guide explains 21 CFR Part 11 in plain words; it is not legal advice. Check the current official text before acting on it, and remember that compliance depends on how a system is validated and used, not on the software alone. Last updated .