21 CFR Part 11, explainedElectronic records and signatures the FDA will accept

Part 11 is the FDA rule that lets a regulated company keep its records, and sign them, electronically instead of on paper, as long as the system can prove who did what, when, and that nothing was quietly changed. This guide takes each requirement in turn, then covers what inspectors look for and what to ask of software.

Issued by
US FDA, Title 21 of the Code of Federal Regulations, Part 11
Applies to
Electronic records that FDA's record rules require (drug GMP in 21 CFR 211, devices, GLP and others), electronic records submitted to FDA, and the signatures on them. Foreign sites supplying the US are inspected against it
In force
Published 20 March 1997, effective 20 August 1997; enforcement narrowed by FDA guidance in 2003

In plain words

What 21 CFR Part 11 is about

Part 11 does not decide which records you keep. The predicate rules do, for example 21 CFR Part 211 for drug manufacturing. Part 11 decides when an electronic version of those records, and an electronic signature on them, is as trustworthy as paper and ink.

It has three subparts: general provisions (scope, implementation and definitions), electronic records (controls that keep records authentic and intact, so a signer cannot readily disown a signed record as not genuine) and electronic signatures (how a signature is made, linked to its record and protected).

Which records it covers was set out in FDA's 2003 guidance on its scope and application: a required record kept electronically instead of on paper, or kept both ways when the electronic version is the one relied on for regulated work; electronic records submitted to FDA; and electronic signatures meant to replace handwritten ones. FDA recommends deciding in advance, and writing down, which version of each record you rely on.

The same guidance narrowed enforcement. FDA said it would use enforcement discretion on validation, audit trails, record retention and record copying, and on legacy systems that were operating before 20 August 1997, met the predicate rules then, still meet them, and are documented as fit for use. It still enforces the rest of Part 11 and the predicate rules themselves. In practice inspectors still expect validated systems with audit trails for GMP records, because the predicate rules and FDA's data integrity guidance call for them.

What it asks for

The requirements, clause by clause

  1. §11.10(a)

    Validated systems

    Validate the system for accuracy, reliability and consistent intended performance, and for its ability to tell when a record is invalid or has been altered.

  2. §11.10(b)

    Accurate copies

    Produce complete, accurate copies in human-readable and electronic form that FDA can inspect, review and copy.

  3. §11.10(c)

    Protected records

    Protect records so they can be retrieved accurately and readily throughout the retention period.

  4. §11.10(d)

    Limited access

    Limit system access to authorised individuals.

  5. §11.10(e)

    Audit trail

    Keep secure, computer-generated, time-stamped audit trails that independently record the date and time of operator entries and actions that create, modify or delete records. Changes must not obscure what was recorded before. The trail is kept at least as long as the records and is available for FDA to review and copy.

  6. §11.10(f)–(h)

    System, authority and device checks

    Enforce the permitted sequence of steps, check that the user is authorised for each action (signing, altering a record, using an input or output device) and, where needed, check that the data source is valid.

  7. §11.10(i)

    Trained people

    Everyone who develops, maintains or uses the system has the education, training and experience for their tasks.

  8. §11.10(j)

    Accountability policy

    Written policies hold people accountable for actions taken under their electronic signatures, to deter falsification.

  1. §11.10(k)

    System documentation

    Control the distribution of, access to and use of system documentation, with revision and change control that keeps a time-sequenced record of how the documentation was developed and changed.

  2. §11.30

    Open systems

    Where access to a system is not controlled by the people responsible for its records, apply the §11.10 controls as appropriate, plus measures such as document encryption and digital signature standards where needed to keep records authentic, intact and confidential.

  3. §11.50

    What a signature shows

    A signed record shows the signer's printed name, the date and time of signing, and the meaning of the signature, such as review, approval or authorship, on screen and on printouts.

  4. §11.70

    Signature linking

    Electronic signatures, and handwritten signatures on electronic records, are linked to their records so they cannot be removed, copied or otherwise transferred to falsify another record by ordinary means.

  5. §11.100

    One signature, one person

    Each e-signature belongs to one individual and is never reused by or reassigned to anyone else, and identity is verified before it is issued. Before or at first use, the company certifies to FDA, over a handwritten signature, that its e-signatures are the legally binding equivalent of handwritten ones; FDA can ask for more on any one signature.

  6. §11.200

    Two components

    Non-biometric signatures use at least two distinct components, such as a user ID and a password. The first signing in a continuous session uses all of them, later ones at least one that only the owner can execute; signings outside one session use all of them. Only the owner uses a signature, and misusing it would take two or more people colluding.

  7. §11.300

    ID and password controls

    ID and password combinations are unique, checked and revised periodically, and deauthorised when lost or compromised. Safeguards prevent unauthorised use and report attempts immediately to security and, as appropriate, management. Tokens and cards that carry or generate codes are tested at first and periodically.

Where inspections find gaps

Where sites usually slip

  • Shared or generic loginsActions are recorded against "QC1" or "admin", so no entry can be attributed to a person.
  • Audit trails off or unreadThe audit trail is disabled, does not cover GMP-relevant data, or is never reviewed.
  • Users who can erase their own dataAnalysts or operators with administrator rights can delete or overwrite records they generated.
  • Incomplete signaturesSignatures that do not show name, date, time and meaning, or that are not bound to the record they sign.
  • Missing paperworkNo e-signature certification filed with FDA, or no written policy making people accountable for their e-signatures.
  • Unvalidated or uncontrolled systemsThe system was never validated for its intended use, or it has been changed without change control since.

What to ask of software

Questions for any vendor

0 of 7
Ask for a demonstration of each, not a yes.

At a glance

Part 11 and Annex 11 side by side

Most regulated manufacturers that export work to both. The two cover the same ground from different directions: Part 11 is a short US regulation about records and signatures; Annex 11 is an EU guideline about the whole life of a computerised system.

Part 11 and Annex 11 side by side
21 CFR Part 11 (US FDA)EU GMP Annex 11
Legal formFederal regulation (1997), binding in the USGuideline in the EU GMP Guide (2011); also used by PIC/S inspectorates
What it coversElectronic records required by FDA rules, and electronic signaturesEvery computerised system used in GMP activities, across its lifecycle
RiskNarrowed by FDA's 2003 guidance to a risk-based scopeRisk management across the lifecycle is the first section
Audit trailComputer-generated, time-stamped record of create, modify and delete, kept at least as long as the records (§11.10(e))Based on risk; records GMP-relevant changes and deletions with the reason, readable and regularly reviewed (9)
Electronic signatureName, date, time and meaning; linked to the record; two components; a hand-signed certification to FDA (§11.50–11.300)Same effect as a handwritten signature within the company, permanently linked, with date and time (14)
SuppliersNot addressed directlyFormal agreements; supplier assessment and audits based on risk (3)
Keeping it validatedValidation and system documentation control (§11.10(a), (k))Change and configuration management, periodic evaluation, incident management (10, 11, 13)
Continuity and archiveRecords protected and retrievable for the retention period (§11.10(c))Backups with tested restore, business continuity and archiving (7, 16, 17)
StatusRequirements unchanged since 1997; FDA guidance explains how it applies them2011 text in force; a revised draft went to consultation in July 2025 and is not yet final

FAQs

Questions people ask about 21 CFR Part 11

Does 21 CFR Part 11 apply to Indian pharmaceutical manufacturers?

Yes, if you make products for the US market or submit records to FDA. FDA's GMP rules apply to foreign sites that supply the US, and Part 11 applies to the electronic records and signatures those rules require. An Indian plant supplying the US is inspected against it.

Is a printout from a computer system a Part 11 record?

Generally not, when the printout meets the predicate rules and the paper is what people rely on; the computer only produced it. If people rely on the electronic version, FDA may treat that as the record. Part 11 also does not cover paper records that are only sent electronically. And FDA's 2018 data integrity guidance warns that a printout of a dynamic record, such as an FT-IR spectrum that can be reprocessed, is not a true copy of it.

Is software ever "Part 11 certified"?

No. FDA does not certify software, and no vendor can make you compliant. Software can provide the technical controls Part 11 requires; compliance depends on how you validate, configure and use the system, and on your procedures.